Impact of Changing Instrument Settings Parameters
Voyager
Biospectrometry
Workstation User’s Guide 5-49
5
Effects of
adjusting Bin size
A smaller Bin size (lower number of nanoseconds):
Increases resolution between peaks when peaks are
narrow in time width (see Figure 5-11).
Increases the size of the data file because the mass
range is recorded with finer sampling and therefore
increases the number of data points collected.
Figure 5-11 Effect of Bin Size (nsec) on Peak
Resolution and Data Quality
A larger Bin size (higher number of nanoseconds):
Decreases resolution between peaks when peaks are
narrow in time width (see Figure 5-11).
Decreases the size of the data file because the mass
range is recorded with coarser sampling (decreases the
number of data points collected).
Data collected at
2 nanoseconds
Data collected at
8 nanoseconds
Baseline resolution
between 2 adjacent
peaks
Partial resolution
between same
peaks